Handpiece Maintenance

Quick Reference

Handpiece Maintenance
12-2
WARNING
Electrical Shock Hazard - Do not remove cover. Refer servicing to
authorized service personnel.
The following chart lists routine maintenance tasks, when you perform
each task, and the equipment on which you perform them.
When
Daily or when used
Daily or when used
Every 6 months, or
when filter changes
color
After 50 hours of use,
or 200 surgical
procedures,
whichever comes first

Return and Replace the Handpiece

The handpiece has been validated to 50 hours of ultrasonic usage
(i.e. with foot pedal depressed) and 200 procedures. To ensure optimal
performance, it is recommended that you replace the handpiece once it
reaches 50 hours or 200 procedures, whichever comes first. The tracking
of ultrasonic usage hours and the recommendation to replace the
handpiece are to support patients and users by following the
recommendations outlined in the FDA Guidance Document issued
March 17, 2015, "Reprocessing Medical Devices in Health Care Settings:
Validation Methods and Labeling" and by complying with applicable
international standards.
To view the hours of use, with the handpiece connected, go to the
Settings screen (see "Settings Screen" on page 4-3).
Integra recommends that you retain a spare, sterile handpiece for every
surgical procedure.
CUSA
Equipment
Console
Footswitch
Contamination Guard
Handpiece
Clarity Ultrasonic Surgical Aspirator System Operator's Manual
®
Task
Clean and wipe dry.
Clean the footswitch.
Replace the
contamination guard.
Write either the
installation date or the
expiration date on the
guard.
Return handpiece to
Integra and replace
with a new handpiece.