Integra CUSA Clarity Operator's Manual page 69

Ultrasonic surgical aspirator system
Table of Contents

Advertisement

Attaching the Tubing
7-10
6. Push the irrigation tubing into the clip at the base of the handpiece.
NOTE
If needed, the clip can be rotated to minimize strain on the user's
wrist.
NOTE
Handpiece cable is not intended to be inserted into the clip at the base
of the handpiece.
7. Clip the handpiece cable to the aspiration tubing using clips available
along the aspiration tubing.
8. Keep sufficient tubing and cable in the sterile field to allow you to
reposition the console.
9. Pass off the cartridge and smaller loop of the irrigation and aspiration
tubing set and handpiece cable connector outside of the sterile field
for attachment to the console.
CUSA
Clarity Ultrasonic Surgical Aspirator System Operator's Manual
®