Integra CUSA Clarity Operator's Manual page 123

Ultrasonic surgical aspirator system
Table of Contents

Advertisement

Amplitude Troubleshooting
13-4
Condition
Ineffective fragmentation
(due to frequency drift -
temperature dependence)
No fragmentation while using
CUSA Clarity in conjunction
with a monopolar
electrosurgery system
Low or no aspiration
NOTE
To avoid tip blockage during surgery flush with saline solution in
between uses. This removes debris before it can dry and block the tip
while it is not in use.
WARNING
To avoid injury or damage to equipment, place system in standby prior to
changing tip or clearing a clog.
CUSA
®
Recommendation
1. Go into standby.
2. Run handpiece test to update the
frequency optimization.
1. Go into standby.
2. Press Start Prime.
3. Press Stop Prime.
4. Press Start Test.
5. Press Main Screen.
1. Check aspiration at the end of the tip
to make sure the aspiration is being
delivered to the distal end of the tip.
2. Check tubing and connections for
leaks.
3. Verify connections to canister.
4. Utilize the stylet to clear the clog.
Clarity Ultrasonic Surgical Aspirator System Operator's Manual