Cleaning The System; Cleaning The Touchscreen; Cleaning The Console And Cart - Integra CUSA Clarity Operator's Manual

Ultrasonic surgical aspirator system
Table of Contents

Advertisement

Cleaning the System

CUSA
Clarity Ultrasonic Surgical Aspirator System Operator's Manual
®
WARNING
Electric Shock Hazard - Always unplug the CUSA Clarity System
before cleaning it.
CAUTION
Product damage may result if you do not follow the instructions
described in the cleaning section.

Cleaning the Touchscreen

Make sure to the console is unplugged before cleaning it. Clean the touch
screen with a lint-free wipe and a cleaner, such as window/glass cleaner,
eyeglass cleaner, or a quaternary ammonia disinfectant.
NOTE
Do not wipe the touch screen with a sponge because it may scratch the
surface.
NOTE
Never apply cleaner directly to the touch screen.

Cleaning the Console and Cart

1. Unplug the power cord from the wall receptacle and the console.
2. Clean the surface with a lint-free wipe soaked with one of the
following:
• 70% IPA
• Super Sani-Cloth
• Sani-Cloth
Bleach
®
3. Make sure the surfaces remain visibly wet for a minimum of two (2)
minutes, then air dry.
WARNING
Do not allow fluids to enter the console.
®

Cleaning the System

11-11

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents