Sterilization Parameters; Sterilization Tray Packaging - Integra CUSA Clarity Operator's Manual

Ultrasonic surgical aspirator system
Table of Contents

Advertisement

Sterilization Parameters

Sterilization Parameters

11-16
1. Place the handpiece in the appropriate slot
outline on the bottom of the tray.
2. Wrap the handpiece cable around the posts
3. Place the handpiece connector in the appropriate slot
with the outline on the bottom of the tray.
4. Optional. Place the nosecone in the appropriate slot
with the outline on the bottom of the tray.
5. Optional. Place the flue
6. Place the torque base in the appropriate slot
outline on the bottom of the tray.
7. Optional. Place the Extended Use Tips (up to 3) in the appropriate
slots
and align them with the outline on the bottom of the tray.
8. Put on the lid
on the sterilizer tray.
9. Close and latch the lid.
The CUSA Clarity handpiece and components must be sterilized with
steam.

Sterilization Tray Packaging

The handpiece and components can be packaged in the sterilization tray
either:
Wrapped—Sterilization tray double wrapped in hospital CSR material
Flash (Unwrapped)—Sterilization tray unwrapped
The cycles below are acceptable for sterilizing the CUSA Clarity
handpiece and components
Packaging
Temp
132°C (269.6°F)
Wrapped
134°C (273.2°F)
134°C (273.2°F)
Flash
132°C (269.6°F)
(Unwrapped)
CUSA
anywhere on the bottom of the tray.
Type
Prevac
Prevac
Prevac
Prevac
Clarity Ultrasonic Surgical Aspirator System Operator's Manual
®
and align it with the
.
and align it
and align it
and align it with the
Time
Dry Cycle
4 min
30 min
3 min
30 min
18 min
30 min
4 min
None