Ultrasonic Control Scale Adjustments

CUSA
Clarity Ultrasonic Surgical Aspirator System Operator's Manual
®
On the main screen, white ultrasonic control scale adjustment arrows
and the selected values
aspiration, and irrigation settings.

Adjusting the Ultrasonic Control Scale Values

To adjust the ultrasonic control scale values:
1. Select the ultrasonic control scale adjustment arrow.
2. Slide the arrow up to increase the value or down to decrease the
value.
The selected value displays above the control scale.
When the Amplitude pedal is pressed, the amplitude scale border is
highlighted and a feedback bar displays next to the Amplitude control
scale to show that the handpiece is running.
When the Fast Flush pedal is pressed, the Irrigation scale border is
highlighted to show Fast Flush is active.
Control Scale Values
The ultrasonic control scale values are:
Amplitude: scale ranges from 10 to 100% in selectable increments of
5%; default value is 60%.
Tissue Select: scale ranges from Off (0), Low (1), Medium (2), High
(3) to Maximum (4); default value is Off. Maximum (4) provides
maximum selectivity, not maximum power.
Aspiration (Suction): scale ranges from 10 to 100% in selectable
increments of 5%; default value is 60%.