Detailed Test Results; System Ready For Use - Integra CUSA Clarity Operator's Manual

Ultrasonic surgical aspirator system
Table of Contents

Advertisement

System Ready for Use

CUSA
Clarity Ultrasonic Surgical Aspirator System Operator's Manual
®
2. If the test fails:
a. Remove the tip and then re-attach with the torque wrench and
base, and re-run the test. If the error continues to persist, attach
a new tip with the torque wrench and base, and re-run the test.
b. If the error continues to persist after replacing the tip, replace the
entire handpiece.
c. Re-run the test.
3. If the test passes, press the Main Screen button.
After testing the system, a Details button displays. Press this button to
view the following detailed test results.

Detailed Test Results

Tip Frequency Optimization - Finds the best frequency for the tip
Amplitude Test - Measures how close the handpiece is to maximum
amplitude and provides a score from 0 to 100%. The score must be
greater than or equal to 50% to pass the test
Press the Main Screen button to begin surgical use.
For a description of the settings, see "Section 4 Touchscreen Display
and Functions" on page 4-1.
Testing the Handpiece
9-5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents