13.2 Guidance and manufacturer's declaration – electromag-
netic immunity
The ATMOS S 201 Thorax is intended for use in the electromagnetic environment specified
below. The customer or user of the ATMOS S 201 Thorax should ensure that it is used in such an
environment.
Immunity Test
Electrostatic discharge (ESD)
according to IEC 61000-4-2
Fast electrical transient/
burst IEC 61000-4-4
Surges IEC 61000-4-5
Magnetic field at power
frequency 50/60 Hz acc. to
IEC 61000-4-8
Voltage Dips / Dropout IEC
61000-4-11
NOTE UT is the alternating mains voltage prior to application of the test levels.
IEC 60601 Test Level
± 6 kV Contact
± 8 kV Air
± 2 kV Mains
± 1 kV I/Os
1 kV
Differential
2 kV
Common
3 A/m
< 5% UT
(> 95% dip of the UT)
For 0.5 cycles
40% UT
(60% dip of the UT)
For 5 cycles
70% UT
(30 % dip of the UT)
For 25 cycles
< 5% UT
(95% dip of the UT)
For 5 sec
Compliance Level
± 6 kV Contact
± 8 kV Air
± 2 kV Mains
Inapplicable
± 1 kV I/Os
1 kV
Differential
2 kV
Common
applicable
3 A/m
< 5% UT
(> 95% dip of the UT)
For 0.5 cycles
40% UT
(60% dip of the UT)
For 5 cycles
70% UT
(30% dip of the UT)
For 25 cycles
< 5% UT
(95% dip of the UT)
For 5 sec
Electromagnetic
Environment - Guidance
Floors should be wood,
concrete, or ceramics tile.
If floors are synthetic, the
relative humidity should be
at least 30%.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Power frequency magnetic
fields should be that of
a typical commercial or
hospital environment.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of the ATMOS S 201
Thorax requires continued
operation even in case of
interruptions of the energy
supply, it is recommended
that the ATMOS S 201
Thorax be powered by
an uninterruptible power
supply or a battery.
49