4.4.5 Attaching, closing and opening the DDS secretion canister
handle

4.4.6 Attaching/removing the DDS secretion canister

4.4.7 DDS secretion canister hose holder

4.4.8 Inserting the DDS hose adapter

4.4.9 Connecting the suction hose

1. To attach the DDS secretion canister handle,
insert it into the grooves of the canister lid (with
the locking latches open).
2. To close the DDS secretion canister handle, clip
the locking latches under the canister rim. Then
press the clips towards the secretion canister
until they click into place.
3. To open, pull the clips outwards and remove the
locking latches from under the canister rim.
1. To attach the DDS secretion canister, allow it to
slide vertically downwards into the mounting
fixture.
2. To remove the DDS secretion canister, lift it
straight up.
1. If using a DDS secretion canister hose holder,
attach it between the canister lid and the hose
adapter.
1. Insert the DDS hose adapter (Ø 6 or 10mm) into
the 'Patient' opening on the DDS canister lid.
2. Turn it slightly and press it down.
 The adapter can be removed again by turning it
slightly.
 Connect the suction hose to the already inserted
hose adapter.
Operation