Intended Purpose Secretion Canister 2 L (Ref 312.1150.5) And Secretion Canister 2 L Without Water Lock (Ref 312.1140.0) - Atmos S 201 Thorax Operating Instructions Manual

Hide thumbs Also See for S 201 Thorax:
Table of Contents

Advertisement

Warnings:
The product is:
Sterility / specific microbial
state:
Single-use device / reprocessing: • The device is intended for multiple use. The device and
For detailed information on the secretion canister and hose system, please refer to the
separate intended uses.
1.3.2 Intended purpose secretion canister 2 l (REF 312.1150.5) and secretion
canister 2 l without water lock (REF 312.1140.0)
Product name:
Main functions:
Intended use:
The following complications may occur during
cardiothoracic drainage:
• Pain due to irritation of the intercostal nerves
• Injury of lung parenchyma / air leak
• Reexpansion pulmonary oedema
• Effusion retention
• Tension pneumothorax
• Cutaneous/subcutaneous emphysema
active
• The device is used in non-sterile condition.
• A sterile secretion canister or a low-germ secretion
canister can be used
• Hose system is sterile.
part of the accessories are reusable. For information on
reprocessing, cleaning and disinfection, please see the
operating instructions.
• Secretion canister and hose system are single-use
devices.
Secretion canister 2 l – (REF 312.1150.5)
Secretion canister 2 l without water lock (REF 312.1140.0)
The secretion canister passes on the controlled vacuum
generated by the ATMOS S 201 Thorax. Fluids and air are
drained through the secretion hose and collected in the
secretion canister. Via balancing scales, the amount of fluid
in the secretion canister is readable and documentable.
An integrated bacterial and viral filter protects the device
against possible contamination and oversuction. For
protection purposes, the pop-off valve opens in case of
overpressure in the secretion canister.
Cover caps ensure proper closure and disposal.
Applies exclusively to the 2 l secretion canister (REF
312.1150.5):
An optional water lock visualises the patient's leak in
bubbles.
Collection of fluids and air from the thorax.
Balancing of the amount of fluid.
Applies exclusively to the 2 l secretion canister (REF
312.1150.5):
• Optional visualisation of air leak in bubbles
• Water lock as one-way valve when using the secretion
canister and the hose system (i.e., without the device)
for temporary gravity drainage
9