⑧ A message box (Fig. 103) will be displayed. Click "OK" to execute analog output at 4mA.
⑨ When you select "Other scale", a window shown in Fig. 104 is displayed for the adjustment of
4mA and 20mA levels. Here, explanation is given for the case with 4mA adjustment.
Enter the reading of the ammeter currently outputting at "4mA output level" box and click "OK" .
When the reading is sent, the transmitter automatically adjusts the output to 4mA. Verify that the
indicator of the connected measuring instrument is 1V.
If another adjustment is required, execute the same procedure.
⑩ To execute 20mA adjustment, click the"20mA"tab on the window in Fig. 104 and execute
adjustment so that the indicator of the connected ammeter shows 20mA as in the same way as
4mA adjustment.
⑪ By clicking "Cancel", a message box (Fig. 101) is displayed. Click "OK" .
Fig.103
Fig.104
E-025-2-E