Trigger
0
1
2
3 4
5
List count=1
List sequence
Steps
Press [Shift]+[3] (List), make sure
1
the
ON
[Enter] to select ON,then press
[
[Enter] to confirm.
2
Set the current range.
3
Set list step count by number
keys,
confirm.
4
Set the first step's current, press
[Enter] to confirm.
5
Set the first step's rise slope,
press [Enter]to confirm.
6
Set the first step's time, such as
5S. Press Enter to confirm.
7
Set the second step's current,
such as 1A, press [Enter] to
confirm.
8
Set the second step's rise slope,
such as 1A/uS press [Enter] to
confirm.
9
Set the second step's time, such
as 5S. Press [Enter] to confirm.
10
Set repeat times, press [Enter] to
confirm.
11
Select the position to save file,
such as 1, press [Enter] to
confirm.
12
Press [
press
[Enter] to confirm (the
indicator will be light now),press
[Esc].
13
Turn on the DC load, press
[Shift]+[
14
Press any function ksys if you want to quit list mode
If you want to run a list file you've saved, please recall it first.The steps is:
Steps
Copyright © Itech Electronics Co., Ltd.
List count=2
Operation
is flashing,if not,press
] key to select EDIT,press
and
press
[Enter]
] to select ON,
] (Trigger) to trigger.
Operation
Basic Operation
VFD Display
LIST
CALL
EDIT
LIST RANGE=3A
LIST STEP=2(2-84)
to
STEP 01 =1A
STEP 01 =0.1A/US
STEP 01 =5S
STEP 01 =2A
STEP 01 =0.1A/US
STEP 01 =5S
REAPEAT =3
SAVE LIST =1(1-7)
LIST
ON
EDIT
Trig
VFD Display
ON
CALL
36