Group1 means you can save and recall parameters in 11-20 registers.
Group2-Group9 can be concluded in the same manner.
Save and Recall operation
For example: the instrucment works in CC mode, setting value is 1A,
Memory Group is 6. Save "CC 1A" in the 61th register and then recall.
1.
Set the parameters ok. To save the instrument's settings to a register,
press [ Shift ]+ [4] (Save). Enter number [6] and [1].
2.
Then press [Enter], The setting is saved.
3.
To recall the instrument's settings from a register, press [ Shift ]+ [Enter]
(Recall).
4.
Enter [1].Then the setting is recalled.
Saving operation will overwrite any values previously saved in that register.
Recalling operation will light the Enter . You should press [ESC] to escape the recalling state
before setting other parameters.
3.11 VON Function
The DC load can be set to only turn on if the voltage is above a set value(VON
set) under configure menu by pressing [ Shift ]+ [9].There are two types of
VON function:Living and Latch.The following will have detailed description for
the two types.
VON set is used to ensure an electronic system under test will not have power applied
unless the supply voltage is above a certain value.If you have no such testing request, do
not set this value arbitrarily.If your instrument can not work normally, for example, set CC=1A,
after turn on the input while the current is still 0A instead of setting value 1A, then you should
check VON set firstly.If VON set is not 0V, then please modify to 0V.
VON LIVING MODE
In Living mode, when power is applied to the DC load, the voltage must rise
above VON setting before the load draws current from the source.If the voltge
below VON setting on the load's terminals, the load will turn off input.
NOTE
NOTE
Copyright © Itech Electronics Co., Ltd.