ERROR MESSAGES

The error code is displayed on the time display.
E.11
Press the select key to clear the error from the display.
6.1

Error message shortform table

E.00
E.10
E.11
E.12
E.13
E.29
E.30
E.31
E.33
E.34
E.36
E.37
E.38
E.50
E.51
E.52
E.57
E.60
E.61
Technical Manual
ERROR
CODE
Exposure key was released too early during the exposure.
X-ray tube Anode voltage (kV) overshoot.
X-ray tube Anode voltage (kV) dropped suddenly.
X-ray tube cathode filament preheating voltages are not calibrated.
Filament preheating voltage calibration failed.
Membrane keyboard key short-circuited/pressed during the self test or
faulty display board.
kV value does not reach or it exceeds the given value (difference more
than 5%).
X-ray tube Anode current (mA) missing, or not in specified limits.
X-ray tube Filament voltage (V) missing, or outside the range (too low
or too high).
X-ray tube Anode voltage (kV) missing, or below the specified limit.
Too long exposure.
kV feedback signal open circuit or short circuit.
mA feedback signal open circuit or short circuit.
Tube head temperature sensor short circuit.
Tube head temperature sensor open circuit.
Filament voltage feedback not in specified limits
Exposure key pressed during self test.
± 15VDC voltage is out of limits
Communication error between control panel and tube head CPU.
ERROR MESSAGE EXPLANATION