7.3
Adjusting the stiffness of the tube head's horizontal axle
Remove the plug from the tube head's axle and adjust the tightness of the two adjusting
screws evenly (arrows on the figure below).
WARNING!
Do not touch the painted screws
or the earth spring attachment screw.
Figure 10
Technical Manual
MECHANICAL ADJUSTMENTS
ø 2.5 Allen screw