SERVICE MODE
2.11

Specifying the PCB versions

Enter the service mode according to the instruction given in section 2.2 "How to enter/exit
the service mode" on page 12.
Press the parameter adjustment up key until the parameter number 20 appears on the kV
display.
20
The value indicating the type of the PCB versions is shown on the time display. 1 = CPU
3
PCB version C3 (or older) with older cascade PCB, 2 = CPU PCB version C5 with newer
cascade PCB and 3 = CPU PCB version E1, works with both old and new cascade PCB.
The factory default is 3.
Press the Select key until the number starts to blink, and the value can now be changed
with the parameter adjustment keys.
Select key
Accept the new setting by pressing the Select key.
Press the Mode key briefly to exit the service mode.
2.12

Ready-state setting

The Planmeca Intra X-ray unit can be set so that the Ready indicator light will only come
on when the Dimaxis program is ready for the exposure, i.e. the Waiting for exposure
message is on the computer screen.
Enter the service mode according to the instruction given in section 2.2 "How to enter/exit
the service mode" on page 12.
Press the parameter adjustment up key until the parameter number 21 appears on the kV
display.
21
The number indicating the ready-state setting (0 or 1) is shown on the time display.
0
0: Normal operation (factory setting). The Planmeca Intra X-ray unit will go into Ready-
state regardless of PC operation.
18
Planmeca Intra X-ray unit
Parameter adjustment up -key
Parameter adjustment keys
Parameter adjustment up -key