Electrical classification
Mechanical data
Weight
Color
Environmental requirements
Ambient temperature
Humidity
Atmospheric pressure range
External mains fuse recommendation
The recommendation for the external mains fuses are:
No other equipment should be connected to the same fused mains line as the x-ray unit. In
some countries an additional external fault current guard is also required.
Original manufacturer
PLANMECA Oy, Asentajankatu 6, FIN-00880, Helsinki, FINLAND
phone: +358-20-7795 500
1.4

User's statement for Planmeca Intra

Radiation leakage technique factors
The maximum rated peak tube potential is 70 kV and the maximum rated continuous tube
current is 0.53 mA for the maximum rated peak tube potential.
Minimum filtration
The radiation port contains an added 1.0 mm aluminium filtration. The measured half-
value is 0.50 - 0.55 at 70 kV. The measured value corresponds to an aluminium equivalent
of 2.0 mm.
Rated line voltage
100, 110-117, 220-240 V~ ±10%. Line voltage regulation 10%.
Maximum line current
6.1 A at 230 V~, 12.2 A at 115 V~
Technical Manual
units with 100V~ or 115V~ voltage setting:
units with 220-240V~ voltage setting:
GENERAL & TECHNICAL DATA
Class I
Type B
total 23 kg (51 lbs)
tube head 4.2 kg (9.3 lbs) with standard cone
4.5 kg (10 lbs) with long cone
RAL 9016
operating +5°C - +40°C
storage -10°C - +50°C
transport -10°C - +50°C
25% - 75%
700 hPa - 1060 hPa
16A, time lag
10A, time lag