Ni-Daq Calibration Function - National Instruments PCI E Series Programmer's Manual

Register-level programmer manual, multifunction i/o boards for pci bus computers, register-level
Hide thumbs Also See for PCI E Series:
Table of Contents

Advertisement

NI-DAQ Calibration Function

© National Instruments Corporation
The NI-DAQ function called
analog input, analog output, and internal reference on the PCI E Series
boards. Due to the complexity of the actual calibration algorithm, use
Calibrate_E_Series
in the EEPROM. You can write a separate application using
Calibrate_E_Series
calibration constants. Writing such an application allows the normal
application to simply copy the calibration constants from the EEPROM
and write them to the calibration DACs upon board initialization.
Calibrate_E_Series
to calibrate each section and store the results
, which is run only when the board needs new
5-17
Chapter 5
Calibration
can calibrate the
PCI E Series RLPM

Advertisement

Table of Contents
loading