Chapter 5
Calibration
EEPROM
Data Stored at EEPROM
Address
394
CALDAC factory constant
393
CALDAC factory constant
392
CALDAC factory constant
391
CALDAC factory constant
390
Factory calibration temp
371
Start of the four user
calibration sections
(MSB × 256) + LSB = Board Code
1
Board Codes
(1 × 256) + 17 = 267
Calibration DACs
Board
PCI-MIO-16E-1
PCI-MIO-16E-4
PCI-6071E
PCI-MIO-16XE-50
PCI-MIO-16XE-10
PCI E Series RLPM
Table 5-6. PCI-6052E EEPROM Map (Continued)
Address
The calibration DACs are used to adjust the errors in the analog signal paths
for errors such as offset and gain. Table 5-7 shows which type of DAC(s)
each board uses for calibration.
Table 5-7. Type of CALDAC Used on Board
MB88341
(8-bit)
—
—
CALDAC
Data Type
Type
AO 1
MB88341
unipolar
linearity
AO 1
MB88341
unipolar
gain coarse
AO 1
MB88341
unipolar
gain fine
AO 1
MB88341
unipolar
offset
—
—
—
—
DAC8800
(8-bit)
—
—
—
5-14
CALDAC
Address
2
10
6
14
—
—
DAC8043
AD8522
1
(12-bit)
(12-bit)
—
—
—
© National Instruments Corporation
Ser
DacLd
Line
1
1
1
1
1
1
2
—
—
—
—