Rotate the display on the transmitter
(optional)
The display on the transmitter electronics module can be rotated 90° or 180° from the
original position.
Figure 2-21:
A.
B.
C.
D.
E.
Procedure
1.
2.
3.
4.
Installation Manual
Rotate the transmitter clockwise to the desired orientation up to 90°.
Secure the cap screw in place and tighten to 60 lb·in (6.8 N·m).
Display components
Transmitter housing
Sub-bezel
Display module
Display screws
Display cover
If the meter is powered up, power it down.
Turn the display cover counterclockwise to remove it from the main enclosure.
Carefully loosen (and remove if necessary) the semi-captive display screws while
holding the display module in place.
Carefully pull the display module out of the main enclosure until the sub-bezel pin
terminals are disengaged from the display module.
Note
If the display pins come out of the board stack with the display module, remove the pins and
reinstall them.
A
B
C
D
Mounting
D
E
23