5 Maintenance
5.5 Disassembly and assembly of components and consumables
5.6

Calibration

The table below lists the type and frequency of calibrations that can be done on the in-
strument. Refer to PrimeView user documentation and to the individual component User
Manuals and Instructions for descriptions of how to perform these calibrations. The
calibrations are performed from PrimeView by selecting System:Calibrate in System
Control.
Component
pH monitor (if applicable)
Pump (if applicable)
Pressure reading
Conductivity
flow cell
48
WARNING
Before disassembly, check that there is no pressure in the piping
system.
WARNING
After assembly, the piping system must be tested for leakage at
maximum pressure for continued protection against injury risks
due to fluid jets, burst pipes or explosive atmosphere.
How often
Every day.
When required.
When required.
Cell constant
Only necessary if specific conductivity with high
accuracy is measured (Cond_Calib).
Temperature
Must be done when changing the conductivity
flow cell (Temp).
Entering a
Must be done when changing the conductivity
new cell con-
flow cell (Cond_Cell).
stant
ÄKTAprime plus Operating Instructions 28-9597-89 AB