4 Operation
4.5 Procedures after a run
Buffers not containing any salt can be left in the system for a short time after a run, even
overnight (not in the pH electrode, see instructions below).
To flush the flow path:
1 Fill a syringe with five times the sample loop volume of deionized water.
2 Rinse the sample loop by injecting the water through the fill port on the injection
3 Put all used inlet tubings in water.
4 In the Templates menu, select Application Template and then System Wash Method.
5 Select the used inlet ports. Inlets A1 and B will always be washed.
6 Press OK to start the method. The system flow path is now automatically flushed.
For information on cleaning and long-term storage, see Section 5.3 Cleaning, on page 46
and Section 5.7 Storage, on page 49.
42
NOTICE
Do not allow particles to enter the UV flow cell as damage to the
flow cell might occur.
NOTICE
If a buffer containing salt has been used, the flow path must be
flushed with deionized water.