2.3 Emergency procedures

2.3
Emergency procedures
In this section
This section describes how to do an emergency shutdown of the ÄKTAprime plus system.
The section also describes the result in the event of power failure.
Emergency shutdown
In an emergency situation, do as follows to stop the run:
Step
1
2
Power failure
The result of a power failure depends on which unit that is affected.
Power failure to...
ÄKTAprime plus
Computer
26
Action
To pause the run without ending the method, press the Pause button located
at the instrument front.
If required, switch off power to the instrument by pressing the Main power
switch to the 0 position. The run is interrupted immediately.
will result in...
The run is interrupted immediately, in an undefined
state
The data collected up to the time of the power failure
is available in PrimeView
The PrimeView computer shuts down in an undefined
state
The run continues, but data cannot be saved in
PrimeView.
ÄKTAprime plus Operating Instructions 28-9597-89 AB