Program List And Explanations; Messages; Diagnostics - Swann AMI Solicon4 Operator's Manual

Version 6.00 and higher
Table of Contents

Advertisement

AMI Solicon4

Program List and Explanations

9.

1 Messages

1.1 Pending Errors
1.1.5
1.2 Message List
1.2.1

2 Diagnostics

2.1 Identification
2.1.3
2.1.4
2.2 Sensors
2.2.1
2.2.1.5
A-96.250.321 / 191015
Program List and Explanations
Provides the list of active errors with their status (active, acknowl-
edged). If an active error is acknowledged, the alarm relay is active
again. Cleared errors are moved to the Message list.
Shows the error history: Error code, date / time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then
the oldest error is cleared to save the newest error (circular buffer).
In diagnostics mode, the values can only be viewed, not modified.
Designation: Designation of the instrument.
Version: Firmware of instrument (e.g. V6.00-10/15)
Factory Test: Test date of the Instrument, Motherboard and Fron-
tend
Operating Time: Years / Days / Hours / Minutes / Seconds
Cond. Sensor
Current value in µS
Raw value in µS
Cell Constant
Contamination
Cal. History: In this menu the calibration values of the last
calibrations are saved.
Number: Numbers the calibrations in descending order.
Date, Time: Date and time of each calibration.
Cell constant: Shows the cell constant of the sensor in use.
Max. 64 data records are memorized. One process calibration cor-
responds to one data record.
53

Advertisement

Table of Contents
loading

This manual is also suitable for:

Amu solicon4

Table of Contents