AMI Solicon4
Program List and Explanations
5.3.1.5
5.3.1.5.1
5.3.1.5.25
5.3.1.5.35
5.3.1.5.45
5.3.2 and 5.3.3
64
Alarm Concentration: Visible if a Conc. parameter has been se-
lected. TDS is displayed in mg/l, all other parameters are displayed
in %.
Alarm High: If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E003, is displayed in the mes-
sage list.
Range: 0.00%–99.90%
Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E004 is displayed in the message
list.
Range: 0.00%–99.90%
Hysteresis: Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluc-
tuates around the alarm value.
Range.0.00%–99.90%
Delay: Duration, the activation of the alarm relay is retarded after
the measuring value has risen above or fallen below the pro-
grammed alarm.
Range: 0–28'800 Sec
Relay 1 and 2: The contacts can be set as normally open or nor-
mally closed with a jumper. See
function of relay contacts 1 or 2 are defined by the user.
Note: The navigation in the menu <Relay 1> and <Relay 2> is
identical. For reason of simplicity only the menu numbers of
Relay 1 are used in the following.
1 First select the functions as:
- Limit upper/lower,
- Control upwards/downwards,
- Timer
- Fieldbus
2 Then enter the necessary data depending on the selected func-
tion.
Relay Contacts 1 and 2, p.
A-96.250.321 / 191015
30. The