Installation - Swann AMI Solicon4 Operator's Manual

Version 6.00 and higher
Table of Contents

Advertisement

AMI Solicon4
Program List and Explanations

5 Installation

5.1 Sensors
5.1.1
5.1.1.1
5.1.1.2
5.1.1.3
5.1.2
5.1.2.1
5.1.3
5.1.3.1
5.1.3.2
5.1.4
A-96.250.321 / 191015
Sensor parameters
Cell Constant: Enter the cell constant printed on the sensor label.
Temp. Corr: Enter the temperature correction printed on the sensor
label.
Cable length: Enter the cable length. Set the cable length to 0.0 m if
the sensors are installed in the flow cell on the AMI monitor.
Temp. Compensation
Comp.: Available compensation models are:
 none
 Coefficient
 non-linear DIN
Flow
Flow measurement: Select the type of flow sensor if a flow sensor
is installed.
Possible flow sensors: None; Q-Flow; deltaT.
Slope: If flow measurement is set to deltaT.
The slope value is used to adjust the flow measurement of the
deltaT sensor if the ambient temperature is higher or lower
than 20 °C.
Conc.
The menu <Concentration> (Conc.) allows the additional measure-
ment of a known substance in the sample. The concentration of the
substance is calculated based on the conductivity of any of the fol-
lowing substances. The calculated value is displayed in %. As an
exception, TDS is displayed in mg/l.
 none
 nitric acid
 hydrochloric acid
 sodium chloride
 caustic soda
 sulfuric acid
 salinity
 TDS as NaCl
 TDS