AMI Solicon4

Default Values

10.
Operation:
Sensors:
Relay Contacts
Logger:
Installation:
Sensors
Signal Output 1
Signal Output 2
Alarm Relay:
72
Default Values
Filter Time Const.:..................................................................10 Sec
Hold after Cal.: ..................................................................... 300 sec
Alarm Relay .................................................. same as in Installation
Relay 1/2 ...................................................... same as in Installation
Input.............................................................. same as in Installation
Logger Interval:.............................................................. 30 Minutes
Clear Logger: ................................................................................no
Sensor Parameters; Cell Constant ...............................0.4000 cm
Sensor Parameters; Temp. corr. ...........................................0.00 °C
Sensor Parameters; Cable length............................................0.0 m
Temp. Compensation; Comp. ...................................................none
Flow; Flow measurement..........................................................none
Conc.: .......................................................................................none
Parameter: .................................................................... Conductivity
Current loop: ......................................................................4–20 mA
Function: .................................................................................. linear
Scaling: Range low: ........................................................... 0.000 µS
Scaling: Range high:............................................................ 100 mS
Parameter: ....................................................................Temperature
Current loop: .....................................................................4 –20 mA
Function: .................................................................................. linear
Scaling: Range low: ................................................................0.0 °C
Scaling: Range high:.............................................................50.0 °C
Alarm Conductivity:
Alarm high:........................................................................... 300 mS
Alarm low: ............................................................................ 0.00 µS
Hysteresis: .......................................................................... 1.00 mS
Delay:.......................................................................................5 Sec
Sample Flow (if Flow measurement = yes):
Flow Alarm.................................................................................. yes
Alarm high:........................................................................... 50.0 l/h
Alarm low: .............................................................................. 5.0 l/h
Sample Temp:
Alarm High: ............................................................................120 °C
Alarm Low:.................................................................................0 °C
A-96.250.321 / 191015