Replacing Fuses - Swann AMI Solicon4 Operator's Manual

Version 6.00 and higher
Table of Contents

Advertisement

AMI Solicon4
Maintenance
6.5.
A-96.250.321 / 191015

Replacing Fuses

WARNING
External Voltage.
External supplied devices connected to relay 1 or 2 or to the
alarm relay can cause electrical shocks.
 Make sure that the devices connected to the following con-
tacts are disconnected from the power before resuming in-
stallation.
– relay 1
– relay 2
– alarm relay
When a fuse has blown, find out the cause and fix it before replac-
ing it with a new one.
Use tweezers or needle-nosed pliers to remove the defective fuse.
Use original fuses provided by SWAN only.
A
A
1.6 AT/250V Instrument power supply
B
1.0 AT/250V Relay 1
C
1.0 AT/250V Relay 2
D
1.0 AT/250V Alarm relay
E
1.0 AF/125V Signal output 2
F
1.0 AF/125V Signal output 1
B
C
D E F