Calibration - Swann AMI Solicon4 Operator's Manual

Version 6.00 and higher
Table of Contents

Advertisement

AMI Solicon4
Maintenance
6.4.
Reagent for
the

Calibration:

42
Calibration
Since the sensor is very reliable a calibration is normally not neces-
sary.
A calibration is recommended if:
 the cell constant is not known
 the sensor is contaminated
 the maintenance measurement shows a discrepancy.
Calibration solution 1.413 mS/cm (25 °C) 1000 ml. Prepare accord-
ing to DIN 38404 / ISO 7888: 1985 / EN 27888.
1 Stop the sample flow.
2 Navigate to menu <Maintenance> /<Calibration>.
3 Press [Enter] and follow the dialog on the Display.
4 Remove the sensor from the flow cell.
5 Clean the sensor carefully and rinse it with clean water, see
Maintenance of the Sensor, p.
6 Use a beaker and fill it with one liter calibration solution.
The beakers diameter must be so large that the sensor has a
distance of min. 3 cm from its edge.
7 Put the sensor into the beaker filled with calibration solution.
Calibration
Clean the sensor
and place it in
standard solution
[Enter] to continue
Calibration
Sensor must have a
min. distance of 3 cm
from the beakers edge
[Enter] to continue
Calibration
Standard solution
10.07 S
Current Value
Cell constant
0.406 cm
Progress
41.
8 Wait at least 5 minutes to permit
3.1.5
temperature equilibration between
sensor and calibration solution.
9 Start the calibration procedure.
3.1.5
10 Press [Enter], to save the values if
3.1.1
the calibration was successful.
1.41 mS
11 Install the sensor into the flow cell.
-1
A-96.250.321 / 191015