AMI Solicon4
Product Description
2.4.2
Sample
connections
Sample
conditions
Dimensions
14

Flow Cell M-Flow PG and M-Flow G1

 Flow cell M-Flow PG, made of PVC and acrylic glass with one
13.5 mm sensor opening for Swansensor Shurecon P.
or
 Flow cell M-Flow G1, made of PVC and acrylic glass with one
G 1" sensor opening for Swansensor Shurecon S.
142
90
33
38
10
60
Inlet:
Outlet:
Each equipped with an elbow hose nozzle for 10 mm tube
Only valid for the flow cell without sensor.
Flow rate:
Temperature:
Inlet pressure:
Sample outlet:
Particle size:
No strong acids and bases.
No organic solvents.
Width:
Front-to-back:
Height:
Panel mounting:
9
G 1/4" thread
G 1/4" thread
4 to 15 l/h
up to 50 °C
up to 1 bar @ 25 °C
Pressure-free outlet
below 0.5 mm
90 to 142 mm
105 mm
161 mm
3 screws M5
A-96.250.321 / 191015