Relay Contacts 1 and 2

Note: Rated load 1 AT / 250 VAC
Relay 1 and 2 can be configured as normally open or as normally
closed. Standard for both relays is normally open. To configure a
Relay as normally closed, set the jumper in the upper position.
NOTICE: Some error codes and the instrument status may
influence the status of the relays described below.
Jumper
pos.
Description
Inactive (opened) dur-
ing normal operation
and loss of power.
Active (closed) when a
programmed function is
executed.
Inactive (closed) during
normal operation and
loss of power.
Active (opened) when a
programmed function is
executed.
A
Jumper set as normally open (standard setting)
B
Jumper set as normally closed
For programming see Menu Installation
tions, p.
53.
Relay configuration