AMI Solicon4
Maintenance
6.3.
6.3.1
Cleaning
6.3.2
A-96.250.321 / 191015
Maintenance of the Sensor
Remove the Sensor form the Flow Cell
To remove the sensor form the flow cell proceed as follows:
1 Unscrew and remove the fixing sleeve [A].
2 Pull the conductivity sensor [B] out of the flow cell block [B].
If the sensor is contaminated, take a small brush and clean it with
water and detergents.
In case of heavy contamination with oil or grease, use ethanol to
clean it. Take a soft tissue and clean the tip of the sensor cautious-
ly.
After each cleaning, the sensor must be rinsed with clean water.
Install the Sensor into the Flow Cell
1 Make sure that the washer [C] and the O-ring [D] are in correct
position.
2 Push the sensor through the flow cell bock [E] into the flow cell.
3 Tighten the fixing sleeve [A] to fix the sensor.
A
Fixing sleeve
B
Conductivity sensor
A
C
Teflon washer
D
O-Ring
E
Flow cell block
B
C
D
E
41