The following description assumes that the installation of the deltaT
sensor takes place after commissioning of the monitor.
Install the deltaT sensor in vertical position with the sample inlet [F]
and cable gland looking downwards.
To ensure laminar flow, the sample inlet must not be restricted; e.g.
any fitting which creates turbulences.
A
B
C
D
E
F
A
Panel
B
Tube connection
C
Elbow hose nozzle
D
Hose nozzle at deltaT sensor outlet
E
deltaT sensor
F
Hose nozzle at deltaT sensor inlet
A-96.250.321 / 191015