AMI Solicon4
Product Description
Fluidics
A-96.250.321 / 191015
The flow cell M-Flow PG consists of the flow cell block [B] and the
calibration vessel [D].
The conductivity sensor [A] is fixed in the flow cell block [B] with a
threaded sleeve.
The sample enters at the sample inlet [F]. It flows through the del-
taT flow sensor [C] (if installed) and then through the flow cell block
into the calibration vessel [D], were the conductivity is measured.
The sample leaves the calibration vessel via flow cell block through
the sample outlet [E] and flows into the drain [G].
C
F
A
Conductivity sensor
B
Flow cell block
C
deltaT flow sensor (option)
D
Calibration vessel
A
B
D
E
G
E
Sample outlet
F
Sample inlet
G
Drain
9