Mounting of Instrument Panel

To mount the instrument panel proceed according the following list:
 The instrument must only be installed by trained personnel.
 Mount the instrument in vertical position.
 For ease of operation mount it so that the display is at eye
level.
 For the installation a kit containing the following installation
material is available:
– 6 Screws 6x60 mm
– 6 Dowels
– 6 Washers 6.4/12 mm
The instrument is only intended for indoor installation, dimensions
see
Dimensions, p.

Connecting Sample Inlet and Outlet

Use a plastic tube (FEP, PA, or PE 10x12 mm) to connect the sam-
ple inlet at the flow cell.
A
If a deltaT senor is installed, push the plastic tube over the hose
nozzle of the deltaT sensor.
Connect the flexible tube 20x15 mm [C] to the 1/2" hose nozzle [B]
and place it into a drain with atmospheric pressure.
A
Waste funnel
A
B
1/2" Hose nozzle
C
Flexible tube 20x15 mm
B
C
11.
A
B
B
Plastic tube 10x12
Elbow hose nozzle