AMI Solicon4
Product Description
2.2.
Power Supply
Electronics
housing
Sample
requirements
On-site
requirements
Measuring
Range
Accuracy
10

Instrument Specification

Voltage:
Power consumption:
Aluminium with a protection degree of IP 66 / NEMA 4X
Ambient temperature:
Limit range of operation:
Storage and transport:
Humidity:
Display:
Flow rate:
Temperature:
Inlet pressure:
Outlet pressure:
The analyzer site must permit connections to:
Sample inlet:
Sample outlet:
Max. Altitude:
Range
0.10 to 9.99 S/cm
10.0 to 99.9 S/cm
100 to 999 S/cm
1.00 to 9.99 mS/cm
10.0 to 29.9 mS/cm
30 to 100 mS/cm
± 0.5% of measured value
100–240 VAC (± 10%)
50/60 Hz (± 5%)
or 24 VDC (± 10%)
max. 30 VA
-10 to +50 °C
-25 to +65 °C
-30 to +85 °C
10–90% rel., non condensing
backlit LCD, 75 x 45 mm
4 to 15 l/h
up to 50 °C
up to 1 bar
pressure free
Flow celll with elbow hose nozzle 1/4"
diam.10 mm
or
deltaT sensor with hose nozzle
diam. 10 mm
1/2" hose nozzle for flexible tube
diam. 20 x 15 mm
2000 m above sea level
Resolution
0.01 S/cm
0.1 S/cm
1 S/cm
0.01 mS/cm
0.1 mS/cm
1 mS/cm