AMI Solicon4
Program List and Explanations
5.4 Miscellaneous
5.4.1
5.4.2
5.4.3
5.4.4
5.4.4.1
5.4.4.2
5.4.4.3
5.4.4.4
5.4.5
70
Language: Set the desired language.
Language
German
English
French
Spanish
Set defaults: Reset the instrument to factory default values in three
different ways:
Set defaults
no
Calibration
In parts
Completely
 Calibration: Sets calibration values back to default. All other
values are kept in memory.
 In parts: Communication parameters are kept in memory. All
other values are set back to default values.
 Completely: Sets back all values including communication
parameters.
Load Firmware: Firmware updates should be done by instructed
service personnel only.
Load Firmware
no
yes
Password: Select a password different from 0000 to prevent unau-
thorized access to the following menus:
Messages
Maintenance
Operation
Installation.
Each menu may be protected by a different password.
If you forgot the passwords, contact the closest SWAN representative.
Sample ID: Identify the process value with any meaning full text,
such as KKS number.