Swann AMI Solicon4 Operator's Manual page 64

Version 6.00 and higher
Table of Contents

Advertisement

AMI Solicon4
Program List and Explanations
5.3 Relay Contacts
5.3.1
5.3.1.1
5.3.1.1.1
5.3.1.1.25
5.3.1.1.35
5.3.1.1.45
62
Alarm Relay: The alarm relay is used as cumulative error indicator.
Under normal operating conditions the contact is active.
The contact is inactive at:
 Power loss
 Detection of system faults like defective sensors or electronic
parts
 High case temperature
 Process values out of programmed ranges.
Note: Above definition assumes that the alarm relay is used as
normally closed (Terminals 10/11). If terminals 12/11 are used,
the alarm relay is normally open, then the above definition is
inverted. See also
p. 27
Program alarm levels, hysteresis values and delay times for the fol-
lowing parameters:
 Alarm Conductivity
 Sample Flow
 Sample Temp.
 Case Temp.
 Alarm Concentration (visible if a Conc. parameter has been
selected)
Alarm Conductivity
Alarm High: If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E001, is displayed in the mes-
sage list.
Range: 0 S–300 mS
Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E002 is displayed in the message
list.
Range: 0 S–300 mS
Hysteresis: Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluc-
tuates around the alarm value.
Range. 0 S–300 mS
Delay: Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28'800 Sec
Alarm Relay, p. 29
and
Connection Diagram,
A-96.250.321 / 191015

Advertisement

Table of Contents
loading

This manual is also suitable for:

Amu solicon4

Table of Contents