AMI Solicon4
Program List and Explanations
5.2 Signal Outputs
5.2.1 and 5.2.2
5.2.1.1
5.2.1.2
5.2.1.3
As process
values
58
Signal Output 1 and 2: Assign process value, the current loop
range and a function to each signal output.
Note: The navigation in the menu <Signal Output 1> and
<Signal Output 2> is identical. For reason of simplicity only the
menu numbers of Signal Output 1 are used in the following.
Parameter: Assign one of the process values to the signal output.
Available values:
 Conductivity
 Temperature
 Sample flow
 Cond. uc (uncompensated)
Current Loop: Select the current range of the signal output.
Make sure the connected device works with the same current
range.
Available ranges: 0–20 mA or 4–20 mA
Function: Define if the signal output is used to transmit a process
value or to drive a control unit. Available functions are:
 Linear, bilinear or logarithmic for process values.
See
As process values, p. 58
 Control upwards or control downwards for controllers.
See
As control output, p. 60
The process value can be represented in 3 ways: linear, bilinear or
logarithmic. See graphs below.
[mA]
20
B
10
12
0 / 4
0.0
0.1
0.2
A
linear
B
bilinear
A
0.3
0.4
0.5
X Measured value
X
A-96.250.321 / 191015