Power Supply - Swann AMI Solicon4 Operator's Manual

Version 6.00 and higher
Table of Contents

Advertisement

AMI Solicon4
Installation
3.5.2
Installation
requirements
28

Power Supply

WARNING
Risk of electrical shock
Do not perform any work on electrical components if the trans-
mitter is switched on. Failure to follow safety instructions could
result in serious injury or death.
 Always turn off AC power before manipulating electric parts.
 Installation and maintenance of electrical parts must be per-
formed by professionals
A
B
C
D
A
Power supply connector
B
Neutral conductor, Terminal 2
C
Phase conductor, Terminal 1
D
Protective earth PE
Note: The protective earth wire (Ground) has to be connected
to the grounding terminal.
The installation must meet the following requirements.
 Transmitter mains fuse 1,6 AT
 Mains cable to comply with standards IEC 60227 or IEC
60245; flammable rating FV1
 Mains equipped with an external switch or circuit-breaker
– near the instrument
– easily accessible to the operator
– marked as interrupter for AMI Solicon4
A-96.250.321 / 191015