Swann AMI Solicon4 Operator's Manual page 48

Version 6.00 and higher
Table of Contents

Advertisement

AMI Solicon4
Error List
Error
E001
Cond. Alarm high
E002
Cond. Alarm low
E003
Conc. Alarm high
E004
Conc. Alarm low
E007
Sample Temp. high
E008
Sample Temp. low
E009
Sample Flow high
E010
Sample Flow low
E011
Temp. shorted
E012
Temp. disconnected
E013
Case Temp. high
E014
Case Temp. low
46
Description
Corrective action
– check process
– check programmed value, see
p. 62
– check process
– check programmed value, see
p. 62
– check process
– check programmed value, see
p. 64
– check process
– check programmed value, see
p. 64
– check process
– check programmed value, see
p. 63
– check process
– check programmed value, see
p. 63
– check sample inlet pressure
– check programmed value, see
5.3.1.2.2, p. 63
– check sample inlet pressure
– Check flow regulating valve
– check programmed value, see
5.3.1.2.35, p. 63
– Check wiring of sensor
– Check wiring of sensor
– check case/environment temperature
– check programmed value, see
5.3.1.4.1, p. 63
– check case/environment temperature
– check programmed value, see
5.3.1.4.2, p. 63
5.3.1.1,
5.3.1.1,
5.3.1.5,
5.3.1.5,
5.3.1.3,
5.3.1.3,
A-96.250.321 / 191015