Program List And Explanations; Messages; Diagnostics - Swann AMI Oxytrace Operator's Manual

Hide thumbs Also See for AMI Oxytrace:
Table of Contents

Advertisement

AMI Oxytrace QED

Program List and Explanations

9.

1 Messages

1.1 Pending Errors
1.1.5
1.2 Maintenance List
1.2.5
1.3 Message List
1.3.1

2 Diagnostics

2.1 Identification
2.1.3
2.1.4
2.2 Sensors
2.2.1
A-96.250.531 / 300617
Program List and Explanations
Provides the list of active errors with their status (active, acknowl-
edged). If an active error is acknowledged, the alarm relay is active
again. Cleared errors are moved to the Message list.
Provides the list of necessary maintenance. Cleared maintenance
messages are moved to the Message list.
Shows the error history: Error code, date / time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then
the oldest error is cleared to save the newest error (circular buffer).
In diagnostics mode, the values can only be viewed, not modified.
Desig.: Designation of the instrument.
Version: Firmware of instrument (e.g. 6.20-06/16)
Factory Test: Test date of the Instrument and Motherboard.
Operating Time: Shows the operating time in Years, Days, Hours,
Minutes and Seconds.
Oxytrace G
Current value: Shows the actual measuring value in ppb.
Raw value tc:
Shows the actual temperature compensated mea-
suring value in mA.
Saturation
Shows the actual saturation in %
69

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ami oxytrace qed

Table of Contents