Function = Limit upper/lower:
When the relays are used as upper or lower limit switches, program
the following:
Parameter: select a process value
Setpoint: If the measured value rises above respectively falls below
the set-point, the relay is activated.
Parameter
Range
0 S–300 mS
Conductivity
Temperature
-25 to +270 °C
Sample flow
0–50 l/h
0 S–300 mS
Cond. uc
Hysteresis: within the hysteresis range, the relay does not switch.
This prevents damage of relay contacts when the measured value
fluctuates around the alarm value.
Parameter
Range
0 S–300 mS
Conductivity
Temperature
-25 to +270 °C
Sample flow
0–50 l/h
0 S–300 mS
Cond. uc
Delay: Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range. 0–600 Sec