AMI Solicon4
Instrument Setup
4.
4.1.
4.2.
Programming
A-96.250.321 / 191015
Instrument Setup
Establish Sample Flow
1 Open sample flow tap.
2 Wait until the flow cell is completely filled.
3 Switch on power.
Programming
The sensor characteristics are printed on the label of the sensor.
Program all sensor parameters in Menu 5.1.1.1,
<Installation> <Sensors> <Sensor parameters>:
SW-xx-xx-xx
ZK = 0.0417
SWAN AG
Enter the:
Cell constant [cm
Temperature correction [°C]
Cable length
Note: Cable length [m] Set the cable length to 0.0 m if the
sensors are installed in the flow cell on the AMI monitor.
Program all parameters for external devices (interface, recorders,
etc.). Program all parameters for instrument operation (limits,
alarms). See
Program Overview, S.
gram List and Explanations, S.
Cell constant
Temperature correction
-1
]
48, for explanations, see
53.
Pro-
35