AMI Oxytrace QED
Installation
3.6.
3.7.
3.7.1
Terminals
1)
NC
10/11
Normally
Closed
NO
12/11
Normally
Open
30
Input
NOTICE: Use only potential-free (dry) contacts.
The total resistance (sum of cable resistance and resistance of
the relay contact) must be less than 50 Ω.
Terminals 16 and 42
If the signal output is set to hold, the measurement is interrupted if
input is active.
For programming see Menu Installation
Relay Contacts
Alarm Relay
NOTICE: Max. load 1 A/250 VAC
Alarm output for system errors.
Error codes see
Error List, p.
NOTICE: With certain alarms and certain settings of the AMI
transmitter the alarm relay does not switch. The error, however,
is shown on the display.
Description
Active (opened) during normal
operation.
Inactive (closed) on error and
loss of power.
Active (closed) during normal
operation.
Inactive (opened) on error and
loss of power.
1) usual use
5.3.4, p.
59.
Relay connection
1)
0V
0V
A-96.250.531 / 300617
85.
11
10
12
11
10
12