Power Supply - Swann AMI Oxytrace Operator's Manual

Hide thumbs Also See for AMI Oxytrace:
Table of Contents

Advertisement

AMI Oxytrace QED
Installation
3.5.3
Installation
requirements
A-96.250.531 / 300617

Power Supply

WARNING
Risk of electrical shock
Do not perform any work on electrical components if the trans-
mitter is switched on. Failure to follow safety instructions could
result in serious injury or death.
 Always turn off AC power before manipulating electric parts.
 Installation and maintenance of electrical parts must be per-
formed by professionals.
A
B
C
D
A
Power supply connector
B
Neutral conductor, Terminal 2
C
Phase conductor, Terminal 1
D
Protective earth PE
NOTICE: The protective earth wire (Ground) has to be
connected to the grounding terminal.
The installation must meet the following requirements.
 Mains fuse 1.6 AT
 Mains cable to comply with standards IEC 60227 or IEC
60245; flammable rating FV1
 Mains equipped with an external switch or circuit-breaker
– near the instrument
– easily accessible to the operator
– marked as interrupter for AMI Oxytrace QED
29