Configure process measurement
Overview
Volume Flow Cutoff specifies the lowest volume flow rate that will be reported as
measured. All volume flow rates below this cutoff are reported as 0.
Procedure
Set Volume Flow Cutoff to the value you want to use.
The default value for Volume Flow Cutoff is 0.0 l/sec (liters per second). The lower limit is
0.
4.3
Configure GSV flow measurement
The gas standard volume (GSV) flow measurement parameters control how volume flow is
measured and reported in a gas application.
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time.
Choose one or the other.
4.3.1
Configure Volume Flow Type for gas applications
Display
ProLink III
Field Communicator Configure > Manual Setup > Measurements > GSV > Volume Flow Type > Standard Gas
Fieldbus host
Overview
Volume Flow Type controls whether liquid or gas standard volume flow measurement is
used.
Restriction
Gas standard volume measurement is incompatible with some applications. Set Volume Flow Type
to Liquid if you are using any of the following applications:
•
•
26
Not available
Device Tools > Configuration > Process Measurement > Flow
Volume
Measurement TB > GSV_Gas_Dens (OD Index 62)
Measurement TB > SNS_GSV_FlowUnits (OD Index 67)
Measurement TB > SNS_GSVflowFactor (OD Index 71)
Measurement TB > SNS_GSV_FlowCutoff (OD Index 74)
Petroleum measurement
Concentration measurement
Micro Motion Model 2700 Transmitters with FOUNDATION Fieldbus