Transducer blocks and views
A.8.6
Other additions for concentration measurement
transducer blocks
Table A-25:
Other additions for concentration measurement transducer blocks
Index and name
#56
SNS_ResetEDVolInv
#57
SNS_ResetEDNetMassInv
#58
SNS_ResetEDNetVolInv
#59
SNS_ED_ResetFlag
#60
SNS_ED_EnableDensLow
Extrap
232
Description
Reset concentration measurement
volume inventory
Msg type = METHOD
Data type = Unsigned16 (2)
Store = Static
Access = R/W in any mode
Available in Release 4.0
Reset concentration measurement
net mass inventory
Msg type = METHOD
Data type = Unsigned16 (2)
Store = Static
Access = R/W in any mode
Available in Release 4.0
Reset concentration measurement
net volume inventory
Msg type = METHOD
Data type = Unsigned16 (2)
Store = Static
Access = R/W in any mode
Available in Release 4.0
Reset all enhanced density curve in-
formation
Msg type = METHOD
Data type = Unsigned16 (2)
Store = Static
Access = R/W (OOS)
Available in Release 4.0
Enable low density extrapolation
alarm
Msg type = ENUM
Data type = Unsigned16 (2)
Store = Static
Access = R/W in any mode
Available in Release 4.0
Micro Motion Model 2700 Transmitters with FOUNDATION Fieldbus
List of values
0 = No effect
1 = Reset
0 = No effect
1 = Reset
0 = No effect
1 = Reset
1 = Reset
1 = Enable
HW
700
800
700
800
700
800
800
CDM
FDM
FVM
800
CDM
FDM
FVM