Transducer blocks and views
Table A-16:
Other additions for diagnostics transducer blocks
Index and name
#116
FRF_Max_Current_mA
#117
FRF_KFQ2
#118
SYS_AnalogOutput_Fault
#119
SNS_MV_Failed
#120
ALARM5_STATUS
190
Description
Maximum sensor current (applica-
ble only if meter verification feature
is enabled)
Msg type = VAR
Data type = FLOAT (4)
Store = Static
Access = R/W (OOS)
Available in Release 7.0
KFQ2 linear density correction for
stiffness value
Msg type = VAR
Data type = FLOAT (4)
Store = Static
Access = R/W (OOS)
Available in Release 7.0
Indicates if a critical fault is present
Msg type = VAR
Data type = DS-66 (2)
Store = —
Access = Read only
Available in Release 7.0
Indicates if meter verification failed
Msg type = VAR
Data type = DS-66 (2)
Store = —
Access = Read only
Available in Release 7.0
Status Word 5
Msg type = ENUM
Data type = BIT STRING (2)
Store = Dynamic/20
Access = Read only
Available in Release 8.0
Micro Motion Model 2700 Transmitters with FOUNDATION Fieldbus
(continued)
List of values
Value part of DS-66
0 = No critical fault
1 = Critical fault present
Value part of DS-66
0 = Meter verification did not fail
1 = Meter verification failed
0x0001 = Viscosity out of limits
0x0002 = Sensor check fail
0x0004 = PM: pressure out of
range
0x0008 = Incorrect display type
0x0010 = Incorrect board type
0x0020 = Time period out of
limits
0x0040 = Case/Meter tempera-
ture differential out of range
0x0080 = API non-convergence
0x0100 = Not used
0x0200 = Not used
0x0400 = Not used
0x0800 = Not used
0x1000 = Not used
0x2000 = Not used
0x4000 = Not used
0x8000 = Not used
HW
800
800
700
800
CDM
FDM
FVM
800
CDM
FDM
FVM