Table A-16:
Other additions for diagnostics transducer blocks
Index and name
#64
FRF_AbortCode
#65
FRF_StateAtAbort
#66
FRF_Progress
#67
FRF_StiffOutLimLpo
#68
FRF_StiffOutLimRpo
#69
FRF_StiffnessLpo_mean
Configuration and Use Manual
Description
The reason the meter verification
routine aborted
Msg type = ENUM
Data type = Unsigned16 (2)
Store = Static
Access = Read only
Available in Release 4.0
The state of the meter verification
routine when it aborted
Msg type = VAR
Data type = Unsigned16 (2)
Store = Static
Access = Read only
Available in Release 4.0
Progress (% complete)
Msg type = VAR
Data type = Unsigned16 (2)
Store = Static
Access = Read only
Available in Release 4.0
Is the LPO stiffness out of limits?
Msg type = VAR
Data type = Unsigned16 (2)
Store = Static
Access = Read only
Available in Release 4.0
Is the RPO stiffness out of limits?
Msg type = VAR
Data type = Unsigned16 (2)
Store = Static
Access = Read only
Available in Release 4.0
The current LPO stiffness calculated
as a mean
Msg type = VAR
Data type = FLOAT (4)
Store = Static
Access = Read only
Available in Release 4.0
(continued)
List of values
0 = No error
1 = Manual abort
2 = Watchdog timeout
3 = Frequency drift
4 = High peak drive voltage
5 = High drive current standard
deviation
6 = High drive current mean val-
ue
7 = Drive loop reported error
8 = High delta T standard devia-
tion
9 = High delta T value
10 = State running
1 through 18
Transducer blocks and views
HW
800
800
800
800
800
800
181